![R&S RTO6 Скачать руководство пользователя страница 1346](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461346.webp)
Remote control commands
R&S
®
RTO6
1346
User Manual 1801.6687.02 ─ 05
PROBe<m>:SETup:OFFSet:USEautozero
<AutoZeroOffset>
Includes the AutoZero offset in measurement results. The auto zero error is detected
with
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<AutoZeroOffset>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Firmware/software:
Version 2.70
CHANnel<m>:EATScale
<ExtAttScl>
Sets the attenuation scale for an external divider.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAttScl>
LIN | LOG
*RST:
LIN
Usage:
Asynchronous command
CHANnel<m>:EATTenuation
<ExtAtt>
Sets the attenuation of an external voltage divider that is part of the DUT before the
measuring point. The external attenuation is included in the measurement, and the
instrument shows the results that would be measured before the divider.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAtt>
Values and unit depend on the selected scale (
Range:
Linear scale: 1E-3 to 1E+6, logarithmic scale: -60
dB to 120 dB
Increment: 0.01
*RST:
1
Usage:
Asynchronous command
Acquisition and setup