Parameters:
<Level>
Range:
-10 to 10
Increment: 1E-3
*RST:
Lower = -0.1, upper =0.1
Default unit: V
Usage:
Asynchronous command
TRIGger<m>:SLEW:RANGe
<RangeMode>
Defines the time limit for the slew rate in relation to the upper or lower trigger level (see
on page 1405). The time measurement starts when the signal crosses the first trigger
level - the upper or lower level depending on the selected slope - and stops when the
signal crosses the second level.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode>
INSRange | OUTRange | LTHan | GTHan
INSRange
Triggers on pulses inside a given range. The range is defined by
the slew rate
±
delta.
OUTRange
Triggers on pulses outside a given range. The range is defined
by the slew rate
±
delta.
LTHan
Triggers on pulses shorter than the given slew rate.
GTHan
Triggers on pulses longer than the given slew rate.
*RST:
GTHan
Usage:
Asynchronous command
TRIGger<m>:SLEW:RATE
<Time>
For the ranges "Within" and "Outside", the slew rate defines the center of a range
which is defined by the limits "
±
Delta".
For the ranges "Shorter" and "Longer", the slew rate defines the maximum and mini-
mum slew rate limits, respectively. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope (see
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Trigger