![R&S RTO6 Скачать руководство пользователя страница 1428](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461428.webp)
Remote control commands
R&S
®
RTO6
1428
User Manual 1801.6687.02 ─ 05
TRIGger<m>:ECOupling
<TrigLevEvtCpl>
Event coupling of trigger levels: Sets the trigger levels to the values of the indicated
event. Thus, channel 1 has one trigger level for all events, channel 2 has one trigger
level and so on.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<TrigLevEvtCpl>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
TRIGger<m>:SEQuence:DELay
<Delay>
Sets the time the instrument waits after an A-event until it recognizes B-events.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Delay>
Range:
0 to 50
Increment: 1E-12
*RST:
0
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:SEQuence:COUNt
<Events>
Sets the number of B-events to be fulfilled after an A-event. The last B-event causes
the trigger.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Events>
Range:
1 to 2147483647
Increment: 1
*RST:
1
Usage:
Asynchronous command
TRIGger<m>:SEQuence:RESet:EVENt
<EnabRstEvt>
If set to ON, the trigger sequence is restarted by the R-event if the specified number of
B-event does not occur.
Trigger