Remote control commands
R&S
®
RTO6
1387
User Manual 1801.6687.02 ─ 05
HDEFinition:STATe
<State>
Activates the high definition mode of the instrument.
Parameters:
<State>
ON | OFF
ON: high definition mode, up to 16 bit digital resolution
OFF: normal oscilloscope mode
*RST:
OFF
Usage:
Asynchronous command
HDEFinition:BWIDth
<Bandwidth>
Sets the filter bandwidth for the high definition mode.
Parameters:
<Bandwidth>
Range:
10000 to max. 2 GHz, depending on the instru-
ment bandwidth.
Increment: 1000
*RST:
1E+6
Default unit: Hz
Usage:
Asynchronous command
See
on page 134 for bandwidth limits.
HDEFinition:RESolution?
Returns the resulting digital resolution in high definition mode.
Return values:
<Resolution>
Range:
0 to 16
Increment: 0.1
*RST:
0
Default unit: bit
Usage:
Query only
Asynchronous command
FORMat:BORDer
<ByteOrder>
Sets the endianess.
The command is only relevant for raw data export in high definition mode (16 bit word
length).
Parameters:
<ByteOrder>
LSBFirst | MSBFirst
LSB first: little endian, least significant byte first
MSB first: big endian, most significant byte first
*RST:
LSBFirst
Acquisition and setup