![R&S RTO6 Скачать руководство пользователя страница 2087](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462087.webp)
Remote control commands
R&S
®
RTO6
2087
User Manual 1801.6687.02 ─ 05
BUS<m>:HBTO:ATTN
<Attenuation>
Sets the attenuation factor. It is used to de-amplify the reverse signal before subtract-
ing it from the forward signal.
Suffix:
<m>
.
1..4
Parameters:
<Attenuation>
Range:
-100 to 0
Increment: 0.1
*RST:
-26
Default unit: dB
BUS<m>:HBTO:FDAM
<SourceDAminus>
Selects the DA- source of the provided forward single ended signal, if
is set to
SINGle
.
Suffix:
<m>
.
1..4
Parameters:
<SourceDAminus>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST:
C2W1
BUS<m>:HBTO:FDAP
<SourceDAplus>
Selects the DA+ source of the provided forward single ended signal, if
is set to
SINGle
.
Suffix:
<m>
.
1..4
Parameters:
<SourceDAplus>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST:
C1W1
BUS<m>:HBTO:FDIF
<SrcDiff>
Sets the source of the provided forward differential signal, if
is
set to
DIFFerential
.
Suffix:
<m>
.
1..4