TRIGger<m>:SPI:TYPE
<Type>
Selects the trigger type for SPI analysis.
Parameters:
<Type>
SSACtive | TIMeout | MOSI | MISO | MOMI
SSACtive
Start of the message: slave select signal SS changes to the
active state.
TIMeout
Triggers on the next message start after the "Timeout" time.
MOSI
Triggers on a specified data pattern in that is expected on the
MOSI line. Define the pattern with
MISO
Triggers on a specified data pattern in that is expected on the
MISO line. Define the pattern with
MOMI
Triggers on a specified data patterns on the MISO and MISO
lines.
*RST:
SSACtive
Usage:
Asynchronous command
TRIGger<m>:SPI:PALignment
<DataAlignment>
Defines how the specified data pattern is searched.
Parameters:
<DataAlignment>
WORD | BIT
WORD
The pattern is matched only at word boundaries.
BIT
Bit-by bit: the pattern can be at any position in the data word.
*RST:
WORD
Usage:
Asynchronous command