Remote control commands
R&S
®
RTO6
1612
User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
100E-12
Default unit: s
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:RUNT:POLarity
<SearchName>,<Polarity>
SEARch:TRIGger:RUNT:POLarity?
<SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
See
Chapter 24.4.4, "Polarity parameter"
*RST:
POSitive
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:RUNT:RANGe
<SearchName>,<Mode>
SEARch:TRIGger:RUNT:RANGe?
<SearchName>
Selects how the time limit of the runt pulse is defined based on the runt width and delta
(see
on page 1613 and
Parameters:
<Mode>
ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
Triggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "
±
Delta".
OUTSide
Triggers if the runt length is outside a given time range. The
range definition is the same as for "Within" range.
*RST:
ANY
Search