I/Q software interface (Option R&S
RTO6-K11)
R&S
®
RTO6
1063
User Manual 1801.6687.02 ─ 05
Sample rate
Sets the required sample rate of the output I/Q data.
Remote command:
Record length
Sets the required record length of the output I/Q data. The resulting acquisition time of
the I/Q data is:
Acquisition time = Record length / Sample rate
Remote command:
Channel On/Off
Switches the channel signal on or off. The signal icon appears on the signal bar. The
waveform of the last acquisition is displayed in the diagram.
Remote command:
Carrier frequency
Sets the carrier frequency of the modulated RF signal or of the complex signal in IF
range. The minimum carrier frequency is 10
mHz, the maximum frequency corre-
sponds to the input sample rate.
Remote command:
Sideband (real input)
Defines the frequency position of the RF spectrum in the input signal: normal or
inverse. The position is important for correct down-conversion and filtering.
●
"Normal" position of the RF spectrum
f
c
- f
c
x(t)e
j2πf
c
t
- 2f
c
-
f
c
- f
c
- 2f
c
Down-conversion
Filtering
f
c
- f
c
Input signal, normal
●
"Inverse" position of the RF spectrum
f
c
- f
c
x(t)e
j2πf
c
t
- 2f
c
f
c
- f
c
- 2f
c
Down-conversion
Filtering
f
c
- f
c
Input signal, inverse
Remote command:
I/Q sampling setup