Figure 19-5: Graphical presentation of the transient response properties
Table 19-6: Transient response
Result
Description
Rise time
The time needed for the signal to change from 10% to 90% of the specified
signal high.
Overshoot level
The maximum swing level above the signal high.
Settling time
The time elapsed from the trigger event to the time the output enters and
remains within the tube tolerance band.
Peak time
The time needed for the response to reach the first peak of the overshoot.
Delay to trigger
The time needed for the response to reach half of the signal high value, after
the trigger event.
The following remote commands are used for handling the measurement results:
●
POWer:TRANsient:RESult[:ACTual]?
●
19.6.2.2
Configuring transient response
For details of the configuration settings, see
Chapter 19.6.2.3, "Transient response set-
1. Select "Analysis" > "Power".
2. Under "Output", select "Transient Response".
3. Connect one or more voltage probes and the current probe to the oscilloscope.
4. Deskew the probes as described in
Chapter 19.1.1.1, "Auto deskew"
5. Connect the probes to the DUT as shown in the "Channels" tab:
Output