Deembedding (option R&S
RTO6-K121)
R&S
®
RTO6
1202
User Manual 1801.6687.02 ─ 05
To add a component, tap the
+
icon and select the component type: cable, fixture,
adapter, proven probe, proven cable, or others ("Custom"). The DUT component is
always included in the deembedding. Also, a connected R&S
RT
‑
ZM, R&S
RT-ZD or
R&S
RT-ZS probe with Rohde
&
Schwarz interface is automatically recognized and
included in the deembedding.
The remark "bypassed" indicates that the component is disabled and not considered in
the deembedding calculation as long as it is not configured correctly.
Remote command:
Save / recall deembedding settings
Saves the complete deembedding configuration, or loads a saved deembedding con-
figuration. The file format is XML.
To save the deembedding configuration, enter a filename and tap "Save". Alternatively,
tap "Save as" to open the file selection dialog box, see
To load a deembedding configuration, tap "Open" and select the required file.
Remote command:
Preset Deembedding
Sets the complete deembedding configuration to the default values.
Remote command:
20.2.2
Common settings for components
The "Components" tab has several subtabs to configure each component of the mea-
surement setup that was defined on the "Setup" tab.
Settings that are available on all subtabs are described in this chapter. Other settings
are component-specific and explained in the following chapters.
Deembedding settings