Remote control commands
R&S
®
RTO6
2560
User Manual 1801.6687.02 ─ 05
Example:
See
Chapter 24.3.6.1, "Adding cables to the measurement
Usage:
Setting only
Asynchronous command
DEEMbedding<m>:REMove
<Index>
Deletes the component with the indicated index from the measurement setup.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Setting parameters:
<Index>
Defines the place of the component in the measurement path.
The DUT is always the first component (index = 1), the probe or
input is the last component.
Usage:
Setting only
Asynchronous command
DEEMbedding<m>:CCOunt?
Returns the total number of components in the measurement setup, including disabled
components.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Return values:
<Value>
Integer value, number of components
Usage:
Query only
DEEMbedding<m>:BANDwidth
<Bandwidth>
Sets the maximum bandwidth until which the signal is corrected. This maximum value
is the minimum bandwidth value of probe, probe tip and oscilloscope bandwidth, and
can not be higher than the highest frequency in a used S-parameter file.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Parameters:
<Bandwidth>
Bandwidth limit for correction.
Range:
1E+9 to 16E+9
Increment: 10
*RST:
4E+9
Default unit: Hz
Deembedding (option R&S
RTO6-K121)