Acquisition and waveform setup
R&S
®
RTO6
184
User Manual 1801.6687.02 ─ 05
Scale coupling
If enabled, the vertical scales of P, N, differential and common mode waveforms are
coupled. The scales are related as follows:
VertScale
P
= VertScale
N
= VertScale
CM
= VertScale
Diff
/2
You can disable the scale coupling and set an individual scale for each waveform.
Remote command:
Offset
Sets the offset of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Remote command:
DIFFerential<m>:DIFFerential:OFFSet
Position
Sets the vertical position of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Remote command:
DIFFerential<m>:COMMon:POSition
DIFFerential<m>:DIFFerential:POSition
5.8.3
Deembedding of differential signals
To remove the parasitic effects of the measurement setup from the measured signal,
you can use deembedding for differential signals. Deembedding requires option
R&S
RTO6-K121.
The deembedding of differential signals is represented by one or more 4-port input
components. You can activate deskewing, and set a skew time.
See also:
Chapter 20, "Deembedding (option R&S
1. On the "Vertical" menu, select "Differential Signals".
2. Select the input channels of the positive and negative signal.
3. Enable the differential signal.
4. Select "Deembedding".
5. Select the channel subtab of the positive input.
6. On the "Setup" tab, enable deembedding for the channel.
7. On the "Input" component, select "Configure".
8. If necessary, activate "Use skew offset". Use the deskewing if the P and N wave-
forms are not correctly aligned.
Differential signals