Remote control commands
R&S
®
RTO6
1381
User Manual 1801.6687.02 ─ 05
DIFFerential<m>:COMMon:STATe
<State>
DIFFerential<m>:DIFFerential:STATe
<State>
DIFFerential<m>:NSIGnal:STATe
<State>
DIFFerential<m>:PSIGnal:STATe
<State>
The commands enable the output waveforms if deembedding by software is active. In
this case, all output waveforms can be displayed at the same time, and each output is
enabled separately.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
DIFFerential<m>:COMMon:SCALe
<Scale>
DIFFerential<m>:DIFFerential:SCALe
<Scale>
Set the vertical scale of differential and common mode waveforms, respectively.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Scale>
Scale value, given in Volts per division.
Range:
1E-3 to 100
Increment: 1E-3
*RST:
0.05
Default unit: V/div
Usage:
Asynchronous command
DIFFerential<m>:COUPling
<ScaleCoupling>
If enabled for a differential signal, the vertical scales of P, N, differential and common
mode waveforms are coupled. The scales are related as follows:
VertScale
P
= VertScale
N
= VertScale
CM
= VertScale
Diff
/2
You can disable the scale coupling and set an individual scale for each waveform.
Suffix:
<m>
.
1..2
Differential signal
Acquisition and setup