Remote control commands
R&S
®
RTO6
1514
User Manual 1801.6687.02 ─ 05
Parameters:
<NDbDown>
Range:
0 to 100
Increment: 1
*RST:
20
Default unit: dB
Usage:
Asynchronous command
MEASurement<m>:SPECtrum:PEXCursion
<Value>
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered as a peak. To avoid identifying noise peaks, enter a peak
excursion value that is higher than the noise levels.
Suffix:
<m>
.
1..10
The suffix is irrelevant.
Parameters:
<Value>
Usage:
Asynchronous command
Firmware/software:
Version 2.70
MEASurement<m>:SPECtrum:ATHReshold
<Value>
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<Value>
numeric value
Default unit: dBm
Usage:
Asynchronous command
MEASurement<m>:SPECtrum:RESult<n>:COUNt
<MaxNoOfResults>
Sets the maximum number of measurement results that are listed in the result table.
Available for peak list and harmonic search measurements (
MEASurement:MAIN
is
set to
PLISt
or
HAR
).
Automatic measurements