Remote control commands
R&S
®
RTO6
1903
User Manual 1801.6687.02 ─ 05
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
SBUS
.
●
The sources of the serial bus are channel signals: use
BUS<m>:...:SOURce
commands.
●
Decoding is enabled:
ON
.
.............................................................................................1903
........................................................................................... 1904
.......................................................................................... 1904
............................................................................................ 1905
............................................................................................1905
............................................................................................ 1905
........................................................................................... 1905
.........................................................................................1906
..............................................................................................1906
............................................................................................. 1906
............................................................................................. 1906
............................................................................................ 1907
............................................................................................1907
............................................................................................1907
.....................................................................................1907
.................................................................................... 1908
................................................................................. 1908
....................................................................................... 1908
...................................................................................... 1909
TRIGger<m>:SENT:TYPE
<Type>
Selects the trigger event for the SENT transmission type.
Parameters:
<Type>
CALI | TSEQ | SMSG | ERRC
CALI
CALIbration: triggers on the falling edge of the calibration/
synchronization pulse.
TSEQ
Transmission SEQuence: triggers either on the falling edge of
the status nibble, or on the last data nibble.
To set the transmission sequence conditions, use
.
To set the data condition, use
,
and
Protocols