Remote control commands
R&S
®
RTO6
1518
User Manual 1801.6687.02 ─ 05
Parameters:
<UpperLimit>
Range:
-100 to 100
Increment: 10E-6
*RST:
0
Usage:
Asynchronous command
MEASurement<m>:SPECtrum:LCHeck<n>:LOWer:MARGin
<LowerMargin>
MEASurement<m>:SPECtrum:LCHeck<n>:UPPer:MARGin
<UpperMargin>
Define the lower and upper margins for the limit check, respectively. Margins are not as
strict as limits and must be within the valid value range. The valid range is defined
using the
MEASurement<m>:SPECtrum:LCHeck<n>:VALid
command.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..15
Number of the spectrummeasurement, see
.
Parameters:
<UpperMargin>
Range:
-100 to 100
Increment: 10E-6
*RST:
0
Usage:
Asynchronous command
24.12.7
Optical measurements
The following table lists the <MeasType> parameter values with a short description.
In addition, the suffixes <n> for limit-check remote commands are listed.
Table 24-9: Optical measurements
<MeasType>
parameter value
(remote control)
Measurement (man-
ual operation)
Description/Result
Suffix
<n>
OAP
Average power
True average component of an optical signal.
1
OMA
Modulation amplitude
Measure of the difference between the optical
power of a non-return-to-zero (NRZ) one pulse and
the optical power of an NRZ zero pulse.
OMA requires an NRZ pattern.
2
MEASurement<m>:OPTical:OMAMplitude:RELative:STARt
..............................................1519
MEASurement<m>:OPTical:OMAMplitude:RELative:STOP
.............................................. 1519
..................................................................................1519
Automatic measurements