Remote control commands
R&S
®
RTO6
1517
User Manual 1801.6687.02 ─ 05
Parameters for setting and query:
<MeasType>
PLISt
Only for peaklist measurements. For other measurements, this
command returns an error.
Example:
Display labels.
MEAS:RES:SHL PLIS,ON
MEASurement<m>:SPECtrum:LCHeck<n>:VALid
<ValidRange>
Enables or disables limit checking for spectrum measurements in the specified mea-
surement channel.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..15
Number of the spectrum measurement, see
.
Parameters:
<ValidRange>
ILIMit | ULIMit | LLIMit | OLIMit
ILIMit
Inside (within) limit; between the upper and lower limit values
ULIMit
Upper limit; above the upper limit value
LLIMit
Lower limit; below the lower limit value
OLIMit
Outside limit; above the upper limit or below the lower limit val-
ues
*RST:
ILIMit
Usage:
Asynchronous command
MEASurement<m>:SPECtrum:LCHeck<n>:LOWer:LIMit
<LowerLimit>
MEASurement<m>:SPECtrum:LCHeck<n>:UPPer:LIMit
<UpperLimit>
Define the lower and upper limits for the limit check, respectively. The valid range is
defined using the
MEASurement<m>:SPECtrum:LCHeck<n>:VALid
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..15
Number of the spectrum measurement, see
.
Automatic measurements