Remote control commands
R&S
®
RTO6
1392
User Manual 1801.6687.02 ─ 05
Parameters:
<Level>
Voltage for the trigger level.
Range:
Depends on vertical scale, channel offset and other
settings. The trigger level must be within the current
display range.
Increment: 1E-3
*RST:
0
Default unit: V
Example:
TRIG:LEV5 0.01
Sets the trigger level for the external trigger signal to 10 mV.
TRIG2:LEV3 0.2
Sets the trigger level for the B-event and B-trigger source chan-
nel 3 to 200 mV.
Usage:
Asynchronous command
TRIGger<m>:FINDlevel
Sets the trigger level automatically. The command is only relevant if the trigger source
is an analog channel CHAN1...4.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Usage:
Event
Asynchronous command
DISPlay:TRIGger:LINes
<State>
Hides or shows the trigger levels in the diagrams.
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
24.9.2
Edge trigger
..........................................................................................1392
TRIGger<m>:EDGE:SLOPe
<Slope>
Defines the edge for the edge trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Trigger