Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1158
User Manual 1801.6687.02 ─ 05
12. If needed adjust the cursors manually.
On the screen, you can see the measurement waveforms of the current and the
voltage. Also, the result box displays the dynamic on resistance. For details, see
Chapter 19.4.2.1, "Dynamic on resistance results"
19.4.2.3
Dynamic on resistance settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 19.2.1, "Channels tab"
In the "Settings" tab, you configure the power measurement parameters and display
settings.
Average
Enables the "Average" method for building the resulting waveform. The average is cal-
culated from the data of the current acquisition and several acquisitions before. The
method reduces random noise and other heterodyne signals. It requires a stable, trig-
gered and periodic signal for correct function. The number of acquisitions for average
calculation is defined with "Average count", and the "Reset mode" defines the restart
condition.
Remote command:
Control loop