![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 286](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864286.webp)
R&S FSP
Instrument Functions
Tracking Generator – Option R&S FSP-B9
4.214
Operating Manual 1164.4556.12 - 06
The permissible setting range is -200 dB to +200 dB in steps of 0.1 dB. Positive off-
sets apply to an amplifier and negative offsets to an attenuator subsequent to the
tracking generator.
The default setting is 0 dB. Offsets <> 0 are marked by the activated enhancement
label
LVL
.
Remote command:
SOUR:POW:OFFS -10dB
4.8.2
Transmission Measurement
This measurement will yield the transmission characteristics of a two-port network.
The built-in tracking generator serves as a signal source. It is connected to the input
connector of the DUT. The input of the R&S FSP is fed from the output of the DUT.
Fig. 4.19 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g.
frequency response of connecting cables).
4.8.2.1
Calibration of Transmission Measurement
SOURCE CAL
The
SOURCE CAL
softkey opens a submenu comprising of the calibration functions
for the transmission and reflection measurement.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...