![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 282](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864282.webp)
R&S FSP
Instrument Functions
TV and RF Trigger – Option R&S FSP-B6
4.210
Operating Manual 1164.4556.12 - 06
The configuration menu for TV trigger settings can be reached with key sequence
TRIG – NEXT – TV TRIG SETTINGS
. The
TRIG
menu is described in section
gering the Sweep – TRIG Key” on page 4.29
.
TRIG
TV TRIG
SETTINGS
The
TV TRIG SETTINGS
softkey switches the TV trigger on and opens a submenu
for configuration of the TV signal parameters.
Remote command:
TRIG:SOUR TV
TV TRIGGER
ON/OFF
The
TV TRIGGER ON/OFF
softkey switches the TV Trigger on or off. When switch-
ing off the TV trigger the selected trigger source will be
FREE RUN.
Remote command:
SENS:TV ON | OFF
VERT SYNC
The
VERT SYNC
softkey configures the trigger for the vertical sync signal. The
R&S FSP triggers on the frame repetition without distinction between the two fields.
Remote command:
TRIG:VID:FIEL:SEL ALL
VERT SYNC
ODD FIELD /
VERT SYNC
EVEN FIELD
The softkeys
VERT SYNC ODD FIELD
and
VERT SYNC EVEN FIELD
configure
the trigger for the vertical sync signal of the first or second field.
Remote command:
TRIG:VID:FIEL:SEL EVEN | ODD
HOR SYNC
The
HOR SYNC
softkey configures the trigger for the horizontal sync signal and
opens the data entry field for selection of the corresponding line.
The trigger can be set on any line number, which can be in the range of 1 to 525 or 1
to 625, depending on the line system. The maximum possible line number will be
selected if the active range is exceeded.
In order to trigger on test line 17 according to CCIR 473-4, for example, the line
value has to be set to 17. This is the default setting after switching on the TV trigger.
Remote command:
TRIG:VID:LINE:NUM 17
TRIG
side menu
!
VERT SYNC ODD FIELD /
VERT SYNC EVEN FIELD
Triggering on TV signals is only possible in time domain (span = 0 Hz). Therefore
the softkey
TV TRIG SETTINGS
is without function in frequency domain.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...