![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 538](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864538.webp)
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.144
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]AVERage[:STATe<1...3>]
ON | OFF
This command switches on or off the average calculation for the selected trace (STATe<1...3>) in the
selected measurement window.
Example:
"AVER OFF"
'Switches off the average calculation for trace 1 in screen A.
"SENS2:AVER:STAT3 ON"
'Switches on the average calculation for trace 3 in screen B.
Characteristics: *RST value:
OFF
SCPI: conform
Mode:
A
[SENSe<1|2>:]AVERage:TYPE
VIDeo | LINear
This command selects the type of average function. If VIDeo is selected, the logarithmic power is
averaged and, if LINear is selected, the power values are averaged before they are converted to
logarithmic values.
The type of average calculation is equally set for all traces in one measurement window.
Example:
"AVER:TYPE LIN"
'Switches screen A to linear average calculation.
Characteristics: *RST value:
VIDeo
SCPI: device-specific
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...