![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 703](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864703.webp)
Operating Manual 1164.4556.12 - 06
6.309
R&S FSP
Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
FSE
[SENSe<1|2>:]AVERage:AUTO
ON | OFF
not available for
R&S FSP
R&S
FSP
FSE
[SENSe<1|2>:]AVERage:COUNt
0 to 32767
FSE
[SENSe<1|2>:]AVERage:COUNt
0 to 32767
R&S
FSP
[SENSe<1|2>:]AVERage:TYPE
VIDeo | LINear
command is used to
select logarithmic or
linear averaging on the
R&S FSP;
therefore parameters are
incompatible to the FSE
FSE
[SENSe<1|2>:]AVERage:TYPE
MAXimum | MINimum |
SCALar
command is used to
select logarithmic or
linear averaging on the
R&S FSP;
therefore parameters are
incompatible to the FSE
R&S
FSP
FSE
[SENSe<1|2>:]AVERage[:STATe<1...3>]
ON | OFF
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:DEMod
<numeric_value>
not available for
R&S FSP and FSET
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:PLL
AUTO | HIGH | MEDium
| LOW
not available for
R&S FSP
R&S
FSP
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo
1Hz to 10MHz
FSET
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo
1Hz to 500MHz
R&S
FSP
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:AUTO
ON | OFF
FSET
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:
EXTernal[:STATe]
ON | OFF
not available for
R&S FSP
R&S
FSP
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio
0.0001 to 1
only numeric values
available. Parameter
ranges differ between
R&S FSP and FSE
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio
0.001 to 1000 | SINe |
PULSe | NOISe
also text parameters are
available. Parameter
ranges differ between
R&S FSP and FSE
not available for FSET
R&S
FSP
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]
10Hz to 10MHz (anal.
filter) 1Hz to 10MHz
(FFT filter)
FSE: 10Hz to 10MHz
(models 20) 1Hz to
10MHz (models 30)
FSET
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]
10 Hz to 500MHz
R&S
FSP
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
AUTO
ON | OFF
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
MODE
ANALog | DIGital
not available for
R&S FSP and FSET
Devices
Command
(Sheet 30 of 47)
Parameter
Notes
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...