R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.170
Operating Manual 1164.4556.12 - 06
6.16.7
SENSe:LIST Subsystem
The commands of this subsystem are used for measuring the power at a list of frequency points with dif-
ferent device settings. The measurement is always performed in the time domain (span = 0 Hz).
A new trigger event is required for each test point (exception: Trigger FREE RUN).
The results are output as a list in the order of the entered frequency points. The number of results per
test point depends on the number of concurrently active measurements (peak/RMS/average).
Selection of concurrently active measurements and setting of parameters that are constant for the whole
measurement is via a configuration command (
SENSe:LIST:POWer:SET
). This also includes the set-
ting for trigger and gate parameters.
The following setting parameters can be selected independently for each frequency point:
• Analyzer frequency
• Reference level
• RF attenuation of attenuator (only with option B25)
• Resolution filter
• Resolution bandwidth
• Video bandwidth
• Measurement time
• Detector
The number of frequencies is limited to 100 entries.
The commands of this subsystem can be used in two different ways:
1. Instrument setup, measurement and querying of the results in a single command:
With this method, there is the least delay between the measurement and the result output. However,
it requires the control computer to wait for the response from the instrument.
2. Instrument setup and querying of the result list at the end of the measurement:
With this method, the control computer may be used for other activities while the measurement is being
performed. However, more time is needed for synchronization via service request.
Settings that are not directly included in commands of this subsystem can be con-
figured by sending the corresponding commands prior to the
SENSe:LIST
com-
mands.
Please note that changes to the trigger level have to be executed in time domain
(span = 0 Hz) in order to take effect for the
SENSe:LIST
commands.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...