R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.50
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:MARKer<1...4>:MINimum:RIGHt
This command positions the specified marker to the next higher minimum value to the right of the
current value (i.e. in ascending X direction) on the corresponding trace.
Example:
"CALC:MARK2:MIN:RIGH"
'Positions marker 2 in screen A to the next higher minimum value to the right of the
current value.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:PEXCursion
<numeric_value>
This command defines the peak excursion, i.e. the spacing below a trace maximum which must be
attained before a new maximum is recognized, or the spacing above a trace minimum which must be
attained before a new minimum is recognized. The set value is valid for all markers and delta markers.
The unit depends on the selected operating mode.
Example:
"CALC:MARK:PEXC 10dB"
'Defines peak excursion 10 dB.
Characteristics:
*RST value: 6dB
SCPI: device-specific
Mode:
A
The numeric suffix in MARKer<1...4> is irrelevant.
CALCulate<1|2>:MARKer<1...4>[:STATe]
ON | OFF
This command switches on or off the selected marker. If no indication is made, marker 1 is selected
automatically. If marker 2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MARK3 ON"
'Switches marker 3 in screen A on or to marker mode.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:TRACe
1 to 3
This command assigns the selected marker (1...4) to the indicated measurement curve. The
corresponding trace must be active, i.e. its status must be different from "BLANK".
If necessary, the corresponding marker is switched on prior to the assignment.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
If no next higher minimum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...