R&S FSP
Instrument Functions
Analyzer Mode
4.28
Operating Manual 1164.4556.12 - 06
SINGLE
SWEEP
The
SINGLE SWEEP
softkey starts n sweeps after triggering. The number of
sweeps is determined by the
SWEEP COUNT
softkey.
When working in the split-screen mode, the frequency ranges of the two windows
are swept one after the other.
If a trace is swept using
TRACE AVERAGE
or
MAXHOLD,
the value set via the
SWEEP COUNT
softkey determines the number of sweeps. If 0 has been entered,
one sweep is performed.
Remote command:
INIT:CONT OFF
CONTINUE
SGL SWEEP
The
CONTINUE SGL SWEEP
softkey repeats the number of sweeps set under
SWEEP COUNT
, however without first deleting the trace.
This is particularly of interest when using the functions
TRACE AVERAGE
and
MAXHOLD
, if previously recorded measurement results are to be taken into consid-
eration for averaging / maximum search.
If
SGL SWEEP DISP OFF
is active, the screen is switched off also during repeated
sweeps.
Remote command:
INIT:CONM
SWEEPTIME
MANUAL
The
SWEEPTIME MANUAL
softkey activates the window for entering the sweep
time manually (see also
BW
menu).
Remote command:
SWE:TIME 10s
SWEEPTIME
AUTO
The
SWEEPTIME AUTO
softkey activates the automatic selection of the sweep time
as a function of the bandwidth of the resolution and video filters (see also
BW
menu).
Remote command:
SWE:TIME:AUTO ON
SWEEP
COUNT
The
SWEEP COUNT
softkey activates the window for the entry of the number of
sweeps to be performed by R&S FSP after a single sweep has been started. If Trace
Average, Max Hold or Min Hold is activated, this also determines the number of
averaging or maximum search procedures.
Example
[
TRACE1
:
MAX HOLD
]
[
SWEEP
:
SWEEP COUNT
: {10}
ENTER
]
[
SINGLE SWEEP
]
R&S FSP performs the Max Hold function over 10 sweeps.
The permissible range for the sweep count is 0 to 32767. For sweep count = 0 or 1,
one sweep is performed. For trace averaging in the continuous-sweep mode,
R&S FSP performs running averaging over 10 sweeps if sweep count = 0; if sweep
count = 1, no averaging, maxhold or minhold is performed.
The sweep count is valid for all the traces in a diagram.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...