Operating Manual 1164.4556.12 - 06
4.27
R&S FSP
Instrument Functions
Analyzer Mode
Fig. 4.2 Reponses of the linear and the logarithmic average detectors as a function of pulse
repetition frequency (PRF) for pulse-modulated signals with pulse durations of 400 ns
and 1 ms. Resolution bandwidth is 1 MHz.
Remote command:
BAND:VID:TYPE LIN
4.5.5
Sweep Settings – SWEEP Key
The SWEEP key serves for configuring the sweep mode and opens the
SWEEP
softkey menu. In split-screen mode, the entries made are valid for the active window
only.
The
CONTINUOUS SWEEP, SINGLE SWEEP
and
SGL SWEEP DISP OFF
soft-
keys are mutually exclusive selection keys.
SWEEP
CONTINUOUS
SWEEP
The
CONTINUOUS SWEEP
softkey activates the continuous sweep mode, which
means that the sweep takes place continuously according to the trigger mode set.
When working in the split-screen mode and with different settings in the two win-
dows, screen A is swept first, followed by screen B. When the softkey is pressed,
the sweep is restarted.
CONTINUOUS SWEEP
is the default setting of R&S FSP.
Remote command:
INIT:CONT ON
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...