![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 149](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864149.webp)
Operating Manual 1164.4556.12 - 06
4.77
R&S FSP
Instrument Functions
Analyzer Mode
The
TIME DOM POWER
softkey activates the power measurement in the time
domain and opens a submenu for configuration of the power measurement.
The submenu allows selection of the type of power measurement (rms or mean
power), the settings for max hold and averaging as well as the definition of limits.
The power evaluation range can be limited by input of limit values.
POWER
(ON OFF)
The
POWER (ON OFF)
softkey switches the power measurement on and off. When
entering the submenu it is
ON
since the power measurement is already switched on
with the
TIME DOM POWER
softkey in the main menu.
The measurement is performed on the trace on which marker 1 is placed. To evalu-
ate another trace, marker 1 should be set on another trace using the
SELECT
TRACE
softkey in
MARKER
menu (
MKR
).
Remote command:
CALC:MARK:FUNC:SUMM:PPE ON
CALC:MARK:FUNC:SUMM:PPE:RES?
CALC:MARK:FUNC:SUMM:RMS ON
CALC:MARK:FUNC:SUMM:RMS:RES?
CALC:MARK:FUNC:SUMM:MEAN ON
CALC:MARK:FUNC:SUMM:MEAN:RES?
CALC:MARK:FUNC:SUMM:SDEV ON
CALC:MARK:FUNC:SUMM:SDEV:RES?
PEAK
The
PEAK
softkey switches on the calculation of the peak value from the points of
the displayed trace or a segment thereof.
For the maximum peak, the largest peak value obtained since the activation of
MAX
HOLD ON
is displayed.
With
AVERAGE ON
, the peak values of a trace are averaged over several sweeps
and displayed.
The number of sweeps over which the average or the maximum value is calculated
is set with the
NUMBER OF SWEEPS
softkey.
Remote command:
CALC:MARK:FUNC:SUMM:PPE ONCALC:MARK:
FUNC:SUMM:PPE:RES?
RMS
The
RMS
softkey switches on the calculation of the rms value from the points of the
displayed trace or a segment of it.
For the maximum peak, the largest rms value obtained since the activation of
MAX
HOLD ON
is displayed.
With
AVERAGE ON
, the rms values of a trace are averaged over several sweeps
and displayed.
This softkey is only available in time domain (span = 0).
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...