R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.156
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]CORRection:TRANsducer[:STATe]
ON | OFF
This command switches the selected transducer factor on or off. Prior to this command, the command
SENS:CORR:TRAN:SEL
must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
This command is available from firmware version 1.40 or higher.
[SENSe<1|2>:]CORRection:TRANsducer:UNIT
<string>
This command specifies the unit for the selected transducer factor. Prior to this command, the
command
SENS:CORR:TRAN:SEL
must be sent.
Parameter:
<string>::= 'DB' | 'DBM' | 'DBMV' | 'DBUV' | 'DBUV/M' | 'DBUA' | 'DBUA/M' | 'DBPW'
| 'DBPT'
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:UNIT 'DBUV'"
Characteristics:
*RST value: 'DB'
SCPI: device-specific
Mode:
A
This command is available from firmware version 1.40 or higher.
[SENSe<1|2>:]CORRection:TRANsducer:VIEW
ON | OFF
This command switches on the display of the active transducer factor or set. Prior to this command,
the command
SENS:CORR:TRAN:SEL
must be sent.
This command is available from firmware version 1.40 or higher.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:VIEW ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:YIG:TEMPerature:AUTO
ON | OFF
This command switches on or off the automatic correction of the YIG filter frequency drift.
When correction is switched on, it is checked once per minute whether the temperature on the YIG
filter has changed by more than 5K relative to the last instance of correction. If this is the case, the
frequency of the YIG filter is – at the end of the next measurement – adjusted as required for the new
temperature. For time-critical applications, the correction function can be switched off after an
operating period of
≥
30 minutes.
The ON parameter is available only if the MW CONV UNIT module has one of the following
modification states:
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...