![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 264](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864264.webp)
R&S FSP
Instrument Functions
Basic Settings
4.192
Operating Manual 1164.4556.12 - 06
4.6.3.8
Firmware Update
For the latest firmware version refer to the Rohde & Schwarz internet site and down-
load the most up-to-date firmware.
A detailed description about performing the firmware update in the intrument is given
in the release note, provided on this CD. The release note is also downloadable
from the Rohde & Schwarz internet site.
The installation of a new firmware version can be performed using a memory stick or
the built-in disk drive.
The installation program is called in the
SETUP
menu.
FIRMWARE
UPDATE
The
FIRMWARE UPDATE
softkey opens the subdirectory for installing/deinstalling
new firmware versions.
Remote command:
--
FIRMWARE
UPDATE
The
FIRMWARE UPDATE
softkey starts the installation program and leads the user
through the remaining steps of the update.
Remote command:
"SYST:FIRM:UPD 'D:\USER\FWUPDATE'"
RESTORE
FIRMWARE
The
RESTORE FIRMWARE
softkey restores the previous firmware version
Remote command:
--
UPDATE PATH
The
UPDATE PATH
softkey is used to select the drive and directories under which
the archive files for the firmware update are stored.
The firmware update can thus also be performed via network drives or USB memory
sticks/USB-CD-ROM drives.
Remote command:
"SYST:FIRM:UPD 'D:\USER\FWUPDATE'"
APPL SETUP
RECOVERY
The
APPL SETUP RECOVERY
(Application Setup Recovery) softkey controls the
instrument behaviour when changing the active application, e.g from
SPECTRUM
to
FM DEMOD
and back from
FM DEMOD
to
SPECTRUM
.
In the default state
OFF
a few parameters of the current analyzer setting are passed
to the application (e.g. center frequency, level settings) or from the application back
to the analyzer mode.
If
APPL SETUP RECOVERY
is switched
ON
, the settings of the applications are
independent of each other. Leaving the
FM DEMOD
application will restore the pre-
vious state of the
ANALYZER
.
The individual application settings are stored on the internal harddisk.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...