2.5.1.1
Measurement Example 1 – Displaying the AF of an AM signal in the time
domain
Test setup:
Settings on the signal generator (e.g. R&S SMIQ):
Measurement with the R&S FSP:
1. Set the R&S FSP to its default state
➢
Press the
PRESET
key.
The R&S FSP is in its default state.
2. Set the center frequency to 100 MHz and the span to 0 kHz
➢
Press the
FREQ
key and enter
100 MHz
.
➢
Press the
SPAN
key and enter
0 Hz
.
3. Set the reference level to +6 dBm and the display range to linear
➢
Press the
AMPT
key and enter
6 dBm
.
➢
Press the
RANGE LINEAR
softkey.
4. Use the video trigger to trigger on the AF signal in order to obtain a
stationary display
➢
Press the
TRIG
key.
➢
Press the
VIDEO
softkey.
The video trigger level is set to 50% if the instrument is switched on for the first
time. The trigger level is displayed as a horizontal line across the graph. The
R&S FSP displays the 1 kHz AF signal stably in the time domain.
Frequency:
100 MHz
Level:
0 dBm
Modulation:
50 % AM, 1 kHz AF
Signal
generator
FSP
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...