![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 575](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864575.webp)
Operating Manual 1164.4556.12 - 06
6.181
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]MIXer:HARMonic:BAND
A|Q|U|V|E|W|F|D|G|Y|J|USER
This command sets the active band.
Example:
"MIX ON"
"MIX:HARM:BAND E"
Characteristics:
*RST value: U
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer:HARMonic:TYPE
ODD | EVEN | EODD
This command sets the type of harmonic.
Example:
"MIX ON"
"MIX:HARM:TYPE EODD"
Characteristics:
*RST value: EVEN
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer:LOPower
<numeric_value>
This command changes the LO level of the external mixer LO port from 13.0 dBm to 17.0 dBm in 0.1
dB steps. Default value is 15.5 dB.
Example:
"MIX ON"
"MIX ON"
"MIX:LOP 16.0DBM"
Characteristics:
*RST value: 15.5 dBm
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer:LOSS:HIGH
<numeric_value>
This command sets the conversion loss of the mixer for higher harmonics in bands with two harmonics.
Example:
"MIX ON"
"MIX:LOSS:HIGH -14DB"
Characteristics:
*RST value: 0 dB
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...