![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 589](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864589.webp)
Operating Manual 1164.4556.12 - 06
6.195
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]POWer:ACHannel:REFerence:TXCHannel:AUTO
MINimum | MAXimum | LHIGhest |
OFF
This command activates or deactivates the automatic selection of a transmission channel to be used
as a reference channel in relative adjacent channel power measurements.
The transmission channel with the highest power, the transmission channel with the lowest power,
or the transmission channel nearest to the adjacent channels can be defined as a reference channel.
The command is available only for multi carrier channel and adjacent channel power measurements
(
CALC:MARK:FUNC:POW:SEL MCAC
) in the frequency domain (span > 0).
Parameter:
MINimum: Transmission channel with the lowest power
MAXimum: Transmission channel with the highest power
LHIGhest: Lowermost transmission channel for the lower adjacent channels,
uppermost transmission channel for the upper adjacent channels
OFF: Deactivates the automatic transmission channel selection.
Example:
"POW:ACH:REF:TXCH:AUTO MAX"
'The transmission channel with the highest power is used as a reference channel.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:ACHannel:REFerence:TXCHannel:MANual
1 to 12
This command selects a transmission channel to be used as a reference channel in relative adjacent
channel power measurements.
The command is available only for multi carrier channel and adjacent channel power measurements
(
CALC:MARK:FUNC:POW:SEL MCAC
) in the frequency domain (span > 0).
Example:
"POW:ACH:REF:TXCH:MAN 3"
'Transmission channel 3 is used as a reference channel.
Characteristics:
*RST value: 1
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:ACHannel:SPACing:ACHannel
100 Hz to 2000 MHz
This command defines the channel spacing of the adjacent channel to the TX channel. At the same
time, the spacing of alternate adjacent channels 1 to 11 is set to the double or triple etc. of the entered
value.
The command is only available in the frequency domain (span > 0).
Example:
"POW:ACH:SPAC:ACH 33kHz"
'Sets the spacing between the carrier signal and
'- the adjacent channel to 33 kHz
'- the alternate adjacent channel 1 to 66 kHz
'- the alternate adjacent channel 2 to 99 kHz
Characteristics:
*RST value: 14 kHz
SCPI: device-specific
Mode:
A-F
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...