Operating Manual 1164.4556.12 - 06
4.67
R&S FSP
Instrument Functions
Analyzer Mode
The
MARKER DEMOD
softkey switches on the audio demodulator and calls a sub-
menu in which the demodulation mode and the duration of the demodulation can be
selected.
Remote command:
CALC:MARK1:FUNC:DEM ON
MKR DEMOD
(ON OFF)
The
MKR DEMOD (ON OFF)
softkey switches the demodulation on/off.
In the frequency range (span >0), the frequency scan is stopped at the frequency of
the active marker with demodulation switched on – provided that the level is above
the threshold line - and the signal is demodulated during the given stop time.
In the time domain (span = 0) demodulation is continuous, i.e. not only active at the
marker position.
Remote command:
CALC:MARK1:FUNC:DEM ON
AM | FM
The softkeys
AM
and
FM
are selector switches one of which only may be active at a
time. They set the desired demodulation mode FM or AM. Default setting is AM.
Remote command:
CALC:MARK1:FUNC:DEM:SEL AM
CALC:MARK1:FUNC:DEM:SEL FM
SQUELCH
The
SQUELCH
softkey enables the input of a level threshold below which the audi-
ble AF is cut off. The squelch function is associated with the internal trigger function
(
TRIGGER
menu), which will be switched on automatically with the squelch.
Squelch level and trigger level do have the same value.
The default setting for the squelch is off.
Remote command:
CALC:MARK1:FUNC:DEM:SQU ON | OFF
CALC:MARK1:FUNC:DEM:SQU:LEV 80 PCT
MKR STOP
TIME
The
MKR STOP TIME
softkey defines the stop time for demodulation at the
marker(s).
The R&S FSP interrupts the frequency sweep at the marker position and activates
the demodulation for the duration of the stop time (see also
MKR DEMOD ON/OFF
).
In the time domain (span = 0) the demodulation is continuously active irrespective of
the stop time set.
Remote command:
CALC:MARK1:FUNC:DEM:HOLD 3s
CONT DEMOD
The
CONT DEMOD
softkey switches on the continuous demodulation in the fre-
quency domain. If the sweep time is long enough, the set frequency range can be
monitored acoustically.
Remote command:
CALC:MARK1:FUNC:DEM:CONT ON
4.5.11.7
Selecting the Trace
MKR
−>
TRACE
The
MKR->TRACE
softkey sets the active marker to different traces. Only those
traces can be selected which are visible on the screen in the same window.
The function of the softkey is identical to that of the softkey with the same name in
the
MKR->
menu.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...