![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 135](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864135.webp)
Operating Manual 1164.4556.12 - 06
4.63
R&S FSP
Instrument Functions
Analyzer Mode
REF POINT
LEVEL
The
REF POINT LEVEL
softkey activates an entry box for the input of a reference
level other than the reference marker level. The function is identical to that of the
softkey with the same name in the
MARKER
menu (
MKR
).
Remote command:
CALC:DELT1:FUNC:FIX:RPO:Y -10dB
REF POINT LVL
OFFSET
The
REF POINT LVL OFFSET
softkey activates an entry box for the input of an
additional level offset for the phase noise calculation.
This level offset is set to 0 dB on when the
REFERENCE FIXED
or
PHASE NOISE
function is enabled.
Remote command:
CALC:DELT:FUNC:FIX:RPO:Y:OFFS 10dB
REF POINT
FREQUENCY
The
REF POINT FREQUENCY
softkey activates an entry box for the manual input
of a reference frequency for the
REFERENCE FIXED
or
PHASE NOISE
function.
Remote command:
CALC:DELT1:FUNC:FIX:RPO:X 10.7MHz
PEAK SEARCH
The PEAK SEARCH sets the reference point level for delta marker 2 in the selected
measurement window to the peak of the selected trace.
Remote command:
CALC:DELT:FUNC:FIX:RPO:MAX
AUTO PEAK
SEARCH
The AUTO PEAK SEARCH softkey activates an automatic peak search for the refer-
ence fixed marker 1 at the end of each particular sweep.
This function may be used for tracking of a drifting source during a phase noise
measurement. The delta marker 2, which shows the phase noise measurement
result, keeps the delta frequency value. Therefore the phase noise measurement
leads to reliable results in a certain offset although the source is drifting. Only if the
marker 2 is reaching the border of the span, the delta marker value is adjusted to be
within the span. In these cases, choose a larger span.
Remote command:
CALC:DELT:FUNC:PNO:AUTO ON | OFF
Measurement example
The phase noise of a CW signal at 100 MHz with 0 dBm level is to be measured at
800 kHz from the carrier
[
PRESET
]
The R&S FSP is set to the default setting.
[
CENTER
: 100
MHz
]
The center frequency is set to 100 MHz.
[
SPAN
: 2
MHz
]
The span is set to 2 MHz.
[
AMPT
: 0
dBm
]
The reference level is set to 0 dBm.
[
MKR FCTN
]
MARKER 1 is switched on and positioned at the maxi-
mum of the displayed trace.
[
PHASE NOISE
: 800
kHz
]
The phase noise measurement is switched on. The
delta marker is positioned on the main marker and the
measured phase noise value is displayed in the marker
info field. The sample detector is used and the video
bandwidth is set to 3 × RBW. When the phase noise
measurement function is enabled, the entry of the delta
marker frequency is activated. It can be entered
directly.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...