Operating Manual 1164.4556.12 - 06
4.213
R&S FSP
Instrument Functions
Tracking Generator – Option R&S FSP-B9
SOURCE ON/
OFF
The
SOURCE ON/OFF
softkey switches the tracking generator on or off. Default
setting is
OFF
Remote command:
OUTP:STAT ON
SOURCE
POWER
The
SOURCE POWER
softkey allows the tracking generator output power to be
selected.
The output power can be set in 0.1 dB steps from -30 dBm to 0 dBm.
If the tracking generator is off, it is automatically switched on when an output power
value is entered.
The default output power is -20 dBm.
Remote command:
SOUR:POW -20dBm
POWER
OFFSET
The
POWER OFFSET
softkey allows selection of a constant level offset for the
tracking generator.
With this offset for example attenuators or amplifiers at the output connector of the
tracking generator can be taken into account for the displayed output power values
on screen or during data entry.
Additional softkeys are available in the displayed menus for controlling an external
generator if option External Generator Control R&S FSP-B10 is fitted. For detailed
information see section
“External Generator Control – Option R&S FSP-B10” on
• When the tracking generator is switched on the maximum stop frequency is
limited to 3 GHz. This upper limit is automatically reduced by a frequency offset
set up for the tracking generator.
•
• In order to meet the data sheet accuracy for measurements with active tracing
generator the start frequency must be set to
≥
3 × Resolution Bandwidth.
• The minimum sweeptime for measurements with data sheet accuracy is 100 ms
in frequency domain (span > 0 Hz). Selecting a sweeptime below this limit will
result in the sweeptime indicator field SWT being supplied with a red asterisk
and the message UNCAL being displayed.
• FFT filters (
FILTER TYPE FFT
in
BW
menu) are not available when the tracking
generator is active.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...