![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 387](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864387.webp)
Operating Manual 1164.4556.12 - 06
5.25
R&S FSP
Remote Control – Basics
Status Reporting System
5.7.3.6
STATus:QUEStionable:ACPLimit Register
This register comprises information about the observance of limits during adjacent
power measurements. It can be queried with commands
STATus:QUEStionable:
ACPLimit:CONDition?
and
STATus:QUEStionable:ACPLimit[:EVENt]?
.
10
LMARgin
This bit is set if a margin is violated (see also section
“STATus:QUEStionable:LMARgin<1|2>
11
This bit is not used
12
ACPLimit
This bit is set if a limit for the adjacent channel power measurement is violated (see also
section
“STATus:QUEStionable:ACPLimit Register” on page 5.25
)
13 to
14
These bits are not used
15
This bit is always 0.
Bit
No.
Meaning
Bit
No.
Meaning
0
ADJ UPPer FAIL(Screen A)
This bit is set if in screen A. the limit is exceeded in the upper adjacent channel
1
ADJ LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower adjacent channel.
2
ALT1 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 1st alternate channel.
3
ALT1 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 1st alternate channel.
4
ALT2 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 2nd alternate channel.
5
ALT2 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 2nd alternate channel.
6
ALT3 to 11 LOWer/UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in one of the lower or upper alternate channels
3 to 11.
7
not used
8
ADJ UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper adjacent channel.
9
ADJ LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower adjacent channel.
10
ALT1 UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper 1st alternate channel.
11
ALT1 LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower 1st alternate channel.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...