R&S FSP
Instrument Functions
Basic Settings
4.170
Operating Manual 1164.4556.12 - 06
NEW
FACTOR /
EDIT TRD FAC
TOR
The
NEW FACTOR
and
EDIT TRD FACTOR
softkeys give access to the submenu
for editing and generating transducer factors.
Depending on the softkey selected, either the table with the data of the factor
marked (
EDIT TRD FACTOR
softkey) or an empty table (
NEW FACTOR
softkey) is
displayed. This table is empty except for the following entries:
The features of the factor are entered in the header of the table, and the frequency
and the transducer factor are entered in the columns.
During editing, a transducer factor remains stored in the background until the factor
edited is saved with the
SAVE TRD FACTOR
softkey or until the table is closed. A
factor that was edited by mistake can be restored by leaving the entry function.
The softkeys in the
UNIT
submenu of the
AMPT
key cannot be operated if the
transducer is on.
Unit:
dB
Interpolation:
LIN for linear frequency scaling
LOG for logarithmic frequency scaling
Name
Entry
of
name
Unit
Selection
of
unit
Interpolation
Selection
of
interpolation
Comment
Entry
of
comment
FREQUENCY
Entry of frequency of reference values
TDF/dB
Entry of transducer factor.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...