![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 592](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864592.webp)
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.198
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]POWer:TRACe
1 to 3
This command assigns the channel/adjacent channel power measurement to the indicated trace in the
selected measurement window. The corresponding trace must be active, i.e. its state must be different
from blank.
Example:
"POW:TRAC 2"
'Assigns the measurement in screen A to trace 2.
"SENS2:POW:TRAC 3"
'Assigns the measurement in screen B to trace 3.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
The measurement of the occupied bandwidth (OBW) is performed on the trace on
which marker 1 is positioned. To evaluate another trace, marker 1 must be posi-
tioned to another trace with
CALCulate:MARKer:TRACe
.