![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 631](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864631.webp)
Operating Manual 1164.4556.12 - 06
6.237
R&S FSP
Remote Control – Description of Commands
TRACe Subsystem
TRACe<1|2>:IMMediate:LEVel?
This query returns the current Y results of the sweep. During a sweep the last measured value is read
out.
Example:
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT"
'starts a sweep (without waiting for the sweep end!)
"TRAC1:IMM:LEV?"
'queries the level of the last measured measurement point
Characteristics:
*RST value:
SCPI: conform
Mode:
A
TRACe<1|2>:IMMediate:RESult?
This query returns the current X and Y results of the sweep. During a sweep the last measured values
are read out.
Example:
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT"
'starts a sweep (without waiting for the sweep end!)
"TRAC:IMM:RES?"
'queries the X and Y values of the last measured measurement point
Characteristics:
*RST value:
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...