![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 95](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864095.webp)
Operating Manual 1164.4556.12 - 06
4.23
R&S FSP
Instrument Functions
Analyzer Mode
This coupling ratio is recommended whenever the amplitudes of pulsed signals are
to be measured correctly. The IF filter is exclusively responsible for pulse shaping.
No additional evaluation is performed by the video filter.
Remote command:
BAND:VID:RAT 10
This setting is only effective for the
VBW AUTO
selection in the main menu.
RBW / VBW
NOISE [10]
The
RBW/VBW NOISE [10]
softkey sets the following coupling ratio:
video bandwidth = resolution bandwidth/10
At this coupling ratio, noise and pulsed signals are suppressed in the video domain.
For noise signals, the average value is displayed.
Remote command:
BAND:VID:RAT 0.1
This setting is only effective for the
VBW AUTO
selection in the main menu.
RBW / VBW
MANUAL
The
RBW/VBW MANUAL
softkey activates the manual input of the coupling ratio.
The resolution bandwidth/video bandwidth ratio can be set in the range 0.001 to
1000.
Remote command:
BAND:VID:RAT 10
This setting is only effective for the
VBW AUTO
selection in the main menu.
SPAN / RBW
AUTO [50]
The
SPAN/RBW AUTO [50]
softkey sets the following coupling ratio:
resolution bandwidth = span/50
This coupling ratio is the default setting of the R&S FSP
Remote command:
BAND:RAT 0.02
This setting is only effective for the
RBW AUTO
selection in the main menu.
SPAN / RBW
MANUAL
The
SPAN/RBW MANUAL
softkey activates the manual input of the coupling ratio.
The span / resolution bandwidth ratio can be set in the range
1 to 10000.
Remote command:
BAND:RAT 0.1
This setting is only effective for the
RBW AUTO
selection in the main menu.
DEFAULT
COUPLING
The
DEFAULT COUPLING
softkey sets all coupled functions to the default state
(
AUTO)
. In addition, the ratio
RBW/VBW
is set to
SINE [1/3]
and the ratio
SPAN/
RBW
to 50 in the
COUPLING RATIO
submenu (default setting,
COUPLING RATIO
softkey not highlighted).
Remote command:
BAND:AUTO ON
BAND:VID:AUTO ON
SWE:TIME:AUTO ON
FILTER TYPE
The
FILTER TYPE
softkey opens the selection list for different filter types. In the
range up to 30 kHz digital band filters with Gaussian characteristic and filtering with
FFT algorithm can be selected.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...