![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 590](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864590.webp)
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.196
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]POWer:ACHannel:SPACing:ALTernate<1...11>
100 Hz to 2000 MHz
This command defines the spacing between the alternate adjacent channels and the TX channel. If
the spacing to an alternate adjacent channel ALTernate<k> is modified, the spacing to all the following
alternate adjacent channels ALTernate<n> is set to (<n> + 1) / (<k> + 1) times the entered value.
This command is only available in the frequency domain (span > 0).
Example:
"POW:ACH:SPAC:ALT1 100kHz"
'Sets the spacing between TX channel and alternate adjacent channel 1 to 100 kHz
and between TX channel and alternate adjacent channel 2 to 150 kHz.
Characteristics:
*RST value:
40 kHz (ALT1)
60 kHz (ALT2)
80 kHz (ALT3)
100 kHz (ALT4)
120 kHz (ALT5)
140 kHz (ALT6)
160 kHz (ALT7)
180 kHz (ALT8)
200 kHz (ALT9)
220 kHz (ALT10)
240 kHz (ALT11)
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:ACHannel:SPACing:CHANnel<1...11>
100 Hz to 2000 MHz
This command defines the channel spacing of the carriers. At the same time the spacing of carriers
with higher channel number are set to the same value. If the spacing is equal between all carriers it is
sufficient to set the spacing between carrier 1 and 2 with the command
SENS:POW:ACP:SPAC:CHAN1
or
SENS:POW:ACP:SPAC:CHAN
. If the spacing are set in ascending order individual spacing of the
carriers can be set.
The command is available only for measurements in the frequency domain (span > 0).
Example:
"POW:ACH:SPAC:CHAN 25kHz"
'Sets the spacing between all carriers to 25 kHz
"POW:ACH:SPAC:CHAN2 4.8MHz"
'sets the spacing between TX carrier 2 and 3 to 4.8 MHz.
Characteristics:
*RST value: 20 kHz
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:ACHannel:TXCHannel:COUNt
1 to 12
This command selects the number of carrier signals.
The command is available only for multi carrier channel and adjacent channel power measurements
(
CALC:MARK:FUNC:POW:SEL MCAC
) in the frequency domain (span > 0).
Example:
"POW:ACH:TXCH:COUN 3"
Characteristics:
*RST value: 4
SCPI: device-specific
Mode:
A-F
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...