![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 641](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864641.webp)
Operating Manual 1164.4556.12 - 06
6.247
R&S FSP
Remote Control – Description of Commands
TRACe Subsystem
Example:
"TRAC:IQ ON"
'Switches on I/Q data acquisition
"TRAC:IQ:SYNC ON ON"
'Switches on synchronization of I/Q data acquisition to the trigger event
"TRAC:IQ:SET NORM,10MHz,32MHz,EXT,POS,0,2048"
'Reads 2048 I/Q-values starting at the trigger point.
'Filtertype: NORMAL (analog)
'RBW: 10 MHz
'Sample Rate: 32 MHz
'Trigger: External
'Slope: Positive
TRAC:IQ:AVER ON
'
Enables averaging of the I/Q measurement data
TRAC:IQ:AVER:COUN 10
'
Selects averaging over 10 data sets
TRAC:IQ:DATA?
'Starts the measurement and reads out the averaged data.
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A-T
This command is available from firmware version 1.40 or higher.