![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 66](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864066.webp)
R&S FSP
Getting Started
Measurements on Modulated Signals
2.42
Operating Manual 1164.4556.12 - 06
Fig. 2-6
Measuring the relative adjacent channel power on a W-CDMA uplink signal
5. Measuring adjacent channel power with the Fast ACP method.
➢
Press the
CP/ACP CONFIG
!
softkey.
➢
Set
FAST ACP
softkey to
ON
.
➢
Press the ADJUST REF LVL softkey.
The R&S FSP measures the power of the individual channels in the time
domain. A root raised cosine filter with the parameters
α
= 0.22 and chip rate
3.84 Mcps (= receive filter for 3GPP W-CDMA) is used as the channel filter.
Fig. 2-7
Measuring the adjacent channel power of a W-CDMA signal with the Fast
ACP method
With W-CDMA, the R&S FSP’s dynamic range for adjacent channel measure-
ments is limited by the 12-bit A/D converter. The greatest dynamic range is,
therefore, obtained with the IBW method.