![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 516](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864516.webp)
R&S FSP
Remote Control – Description of Commands
INITiate Subsystem
6.122
Operating Manual 1164.4556.12 - 06
INITiate<1|2>:DISPlay
ON | OFF
This command configures the behavior of the display during a single sweep.
The numeric suffix of INITiate is irrelevant with this command.
Parameter:
OFF: the display is switched off during the measurement,
ON: the display is switched on during the measurement.
Example:
"INIT2:CONT OFF"
'Switches to single-sweep mode
"INIT2:DISP OFF"
'Sets the display behavior to
OFF
"INIT2;*WAI"
'Starts the measurement with display switched off.
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
INITiate<1|2>:ESPectrum
This command starts a Spectrum Emission Mask measurement. The measurement can be stopped in
continuous sweep mode with
.
Example:
"SENS:SWE:MODE ESP"
' activates the SEM measurement
"INIT:CONT OFF"
' Set to Single Sweep
"INIT:ESP"
' Starts a SEM measurement
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
INITiate<1|2>[:IMMediate]
The command initiates a new sweep in the indicated measurement window.
With Sweep Count > 0 or Average Count > 0, this means a restart of the indicated number of
measurements. With trace functions MAXHold, MINHold and AVERage, the previous results are reset
on restarting the measurement.
In single-sweep mode, synchronization to the end of the indicated number of measurements can be
achieved with the command *OPC, *OPC? or *WAI. In continuous-sweep mode, synchronization to
the sweep end is not possible since the overall measurement never ends.
Example:
"INIT2:CONT OFF"
'Switches to single-sweep mode.
"DISP:WIND:TRAC:MODE AVER
'Switches on trace averaging.
"SWE:COUN 20"
Setting the sweep counter to 20 sweeps.
"INIT2;*WAI"
'Starts the measurement and waits for the end of the 20 sweeps.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...