R&S FSP
Remote Control – Basics
Status Reporting System
5.26
Operating Manual 1164.4556.12 - 06
5.7.3.7
STATus:QUEStionable:FREQuency Register
This register comprises information about the reference and local oscillator.
It can be queried with commands
STATus:QUEStionable:FREQuency:CONDi-
tion?
and
STATus:QUEStionable:FREQuency[:EVENt]?
.
12
ALT2 UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper 2nd alternate channel.
13
ALT2 LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower 2nd alternate channel.
14
This bit is not used.
15
This bit is always set to 0.
Bit
No.
Meaning
Bit
No.
Meaning
0
OVEN COLD
This bit is set if the reference oscillator has not yet attained its operating temperature. 'OCXO'
will then be displayed.
1
LO UNLocked
(Screen A)
This bit is set if the local oscillator no longer locks. 'LOUNL will then be displayed.
2 to 8
not used
9
LO UNLocked
(Screen B)
This bit is set if the local oscillator no longer locks.' LOUNL' will then be displayed.
10 to
14
not used
15
This bit is always 0.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...