![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 175](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864175.webp)
Operating Manual 1164.4556.12 - 06
4.103
R&S FSP
Instrument Functions
Analyzer Mode
Measurement of Occupied Bandwidth
An important characteristics of a modulated signal is its occupied bandwidth. In a
radio communications system for instance the occupied bandwidth must be limited
to enable distortion-free transmission in adjacent channels. The occupied bandwidth
is defined as the bandwidth containing a defined percentage of the total transmitted
power. A percentage between 10% and 99.9% can be set on the R&S FSP.
OCCUPIED
BANDWIDTH
The
OCCUPIED BANDWIDTH
softkey activates measurement of the occupied
bandwidth according to the current configuration and opens the submenu for config-
uring the measurement. The softkey is available only in frequency domain (span >
0) and is highlighted when the measurement is switched on.
In the spectrum display mode, this measurement determines the bandwidth that
contains a predefined percentage of the power of the displayed frequency range (
%
POWER BANDWIDTH
softkey). The occupied bandwidth is output in the marker dis-
play field and marked on the trace by temporary markers.
OCCUP BW
(ON OFF)
The
OCCUP BW ON/OFF
softkey switches measurement of the occupied band-
width on or off.
Remote command:
CALC:MARK:FUNC:POW:SEL OBW
CALC:MARK:FUNC:POW:RES? OBW
CALC:MARK:FUNC:POW OFF
% POWER
BANDWIDTH
The
% POWER BANDWIDTH
softkey opens the entry of the percentage of power
related to the total power in the displayed frequency range which defines the occu-
pied bandwidth (percentage of total power).
The valid range of values is 10% to 99.9%.
Remote command:
SENS:POW:BWID 99PCT
CHANNEL
BANDWIDTH
The
CHANNEL BANDWIDTH
softkey opens an input window for defining the chan-
nel bandwidth for the transmission channel. For measurements in line with a specific
transmission standard, the bandwidth specified by the standard for the transmission
channel must be entered. The default setting is 14 kHz.
The specified channel bandwidth is used for optimization of the test parameters of
the R&S FSP with
ADJUST SETTINGS
.
Remote command:
SENS:POW:ACH:BWID 14kHz
• The softkey is only available in the frequency domain (span > 0).
• The measurement is performed on the trace with marker 1. In order to evaluate
another trace, marker 1 must be placed on another trace by means of
SELECT
TRACE
in the
MARKER
menu
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...