![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 94](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864094.webp)
R&S FSP
Instrument Functions
Analyzer Mode
4.22
Operating Manual 1164.4556.12 - 06
The
SWEEPTIME AUTO
function is designed for detectors like sample, max peak or
average. For sweeping with the quasipeak detector a manual setting of the sweep
time should be used at any time.
Remote command:
SWE:TIME:AUTO ON
COUPLING
RATIO
The
COUPLING RATIO
softkey opens a submenu for selection of the coupling ratios.
When the default setting is active, i.e. the
COUPLING RATIO
softkey is deactivated
(not highlighted), the ratio span/resolution bandwidth (SPAN/RBW) is 50 (this corre-
sponds to SPAN / RBW AUTO [50]) and the ratio resolution bandwidth/video band-
width (RBW/VBW) is 0.33 (this corresponds to RBW / VBW SINE [1/3]).
If the ratio RBW/VBW or SPAN/RBW is different from the default setting, the
COU-
PLING RATIO
softkey is highlighted.
The softkeys
RBW/VBW...
are selection keys. Only one softkey can be enabled at
any one time. The softkeys are only effective for the
VBW AUTO
selection in the
main menu.
The softkeys
SPAN/RBW...
are also selection keys. They are only effective for the
RBW AUTO
selection in the main menu.
RBW / VBW
SINE [1/3]
The
RBW | VBW SINE [1/3]
softkey sets the following coupling ratio:
video bandwidth = 3 × resolution bandwidth.
This is the default setting for the coupling ratio resolution bandwidth/video band-
width.
This is the coupling ratio recommended if sinusoidal signals are to be measured.
Remote command:
BAND:VID:RAT 3
This setting is only effective for the
VBW AUTO
selection in the main menu.
RBW / VBW
PULSE [0.1]
The
RBW | VBW PULSE [0.1]
softkey sets the following coupling ratio:
video bandwidth = 10 × resolution bandwidth or
video bandwidth = 10 MHz (= max. VBW).
Band A
Band B
Band C/D
Frequency range
< 150 kHz
150 kHz to 30 MHz
> 30 MHz
IF bandwidth
200 Hz
9 kHz
120 kHz
Frequency scan rate
20 s/kHz
200 s/MHz
20 s/MHz
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...