4.73
R&S FSP
Instrument Functions
Analyzer Mode
Fig. 4.8 Examples of level measurement with different settings of PEAK EXCURSION
The following table lists the signals as indicated by marker numbers in the diagram
above, as well as the minimum of the amplitude decrease to both sides of the signal:
With
40 dB peak excursion
,
NEXT PEAK
,
NEXT PEAK RIGHT
and
NEXT PEAK
LEFT
will not find any signal, as the signal level does not decrease by more than 30
dB to either side of any signal.
Order of signals detected:
or
With
20 dB peak excursion
,
NEXT PEAK
and
NEXT PEAK RIGHT
will also detect
signal 2, as the signal level decreases at least by 29.85 dB to either side of this sig-
nal, which is now greater than the peak excursion.
signal #
min. amplitude decrease to both sides of the signal
1
30 dB
2
29.85 dB
3
7 dB
4
7 dB
PEAK:
signal 1
NEXT PEAK:
signal 1 (no further signal detected)
PEAK:
signal 1
NEXT PEAK LEFT:
signal 1 (no further signal detected)
NEXT PEAK RIGHT: signal 1 (no further signal detected)
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...