Operating Manual 1164.4556.12 - 06
2.27
R&S FSP
Getting Started
Measurements on Modulated Signals
2.5.2.1
Measurement Example – Displaying the AF of an FM carrier
Test setup:
Settings on the signal generator (e.g. R&S SMIQ):
Measurement with the R&S FSP:
1. Set the R&S FSP to its default state
➢
Press the
PRESET
key.
The R&S FSP is in its default state.
2. Set the center frequency to 99.64 MHz and the span to 300 kHz.
➢
Press the
FREQ
key and enter
99.64 MHz.
➢
Press the
SPAN
key and enter
300 kHz
.
3. Set a resolution bandwidth of 300 kHz.
➢
Press the
BW
key.
➢
Press the
RES BW MANUAL
softkey and enter
300 kHz
.
4. Set a display range of 20 dB and shift the filter characteristics to the middle
of the display.
➢
Press the
AMPT
key.
➢
Press the
RANGE LOG MANUAL
softkey and enter
20 dB
.
➢
Press the
NEXT
key.
➢
Set the
GRID
softkey to
REL
.
➢
Press the
PREV
softkey.
➢
Using the spinwheel, shift the reference level so that the filter edge intersects
the - 10 dB level line at the center frequency.
The slope of the 300 kHz filter is displayed. This corresponds to the
demodulator characteristics for FM signals with a slope of approx. 5 dB/100
kHz.
Frequency:
100 MHz
Level:
-30 dBm
Modulation:
FM 0 kHz deviation (i.e., FM = off), 1 kHz AF
Signal
generator
FSP
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...